Input/Output
More Information
The table below gives the range for each output.
Auto
Selects the Auto state for the Analog Output menu. In this state, the Analog Output will automatically be
set to the most sensible setting for the current mode or measurement.
If you make a selection manually from the Analog Out menu, this selection will remain in force until you
change it (or re-select Auto), even if you go to a mode or measurement for which the selected output
does not apply.
Preset: This is unaffected by Preset but is set to DAUDio on a "Restore Input/Output
Defaults" or "Restore System Defaults->All
State Saved: Saved in Input/Output State
Readback line: 1-of-N selection [variable]
Backwards Compatibility Notes: Prior to A.04.00, OFF was the default functionality except when in the Analog
Demod application or with Tune and Listen, in which case it was DAUDio,
and there was no selection menu. So for backwards compatibility with earlier
X-Series firmware versions, Auto (:OUTP:ANAL:AUTO ON) will duplicate
the prior behavior.
The DNWB and SANalyzer parameters, which were legal in PSA but perform
no function in the X-Series, are accepted without error.
Initial S/W Revision: A.04.00
Analog Out Nominal Range exc.
(10% overrange)
Scale Factor Notes
Off 0 V
Screen Video 0 – 1 V open circuit 10%/division 8566 compatible
Log Video 0 – 1 V terminated 1/(192.66 dB/V) dB referenced to mixer level, 1V out for –10
dBm at the mixer.
Linear Video 0 – 1 V terminated 100%/V Linear referenced to Ref Level, 1 V out for RF
envelope at the Ref Level.
Demod Audio (varies with test set setting)
Key Path:
Input/Output, Output Config, Analog Out
Remote Command: :OUTPut:ANALog:AUTO OFF|ON|0|1
:OUTPut:ANALog:AUTO?
Example: OUTP:ANAL:AUTO ON
Preset: ON
State Saved: Saved in Input/Output State