Trace/Detector
Smooth Trace Data (Remote Command Only)
Not recommended for new designs.Use the CALCulate:DATA:COMPress command instead.
Smooths the trace according to the number of points specified in :TRACe:MATH:SMOoth:POINts.
There is no equivalent front panel function.
The purpose of this function is to perform a spatial video averaging, as compared to the temporal version
supplied by the video-average command [:SENSe]:AVERage:TYPE VIDeo. The functions of
TRACe:MATH:SMOoth <trace> and [:SENSe]:AVERage:TYPE VIDeo|POWer are not
interchangeable.
Number of Points for Smoothing (Remote Command Only)
Not recommended for new designs. (Will not be supported in future designs.) Use the
CALCulate:DATA:COMPress command instead.
Specifies the number of points that are smoothed. Increasing the number of points increases smoothing
at the cost of decreasing resolution. If the number of points is an even number, then the number of points
is increased by one. If the number of points is larger than the number of sweep points, then the number of
sweep points is used, unless the number of sweep points is even, in which case the number of points is
the sweep points minus one. The number of points smoothed is always an odd number.
Mean Trace Data (Remote Command Only)
Not recommended for new designs. Use the CALCulate:DATA:COMPress command instead.
Remote Command: :TRACe:MATH:SMOoth
TRACE1|TRACE2|TRACE3|TRACE4|TRACE5|TRACE6
Initial S/W Revision: Prior to A.02.00
Remote Command: :TRACe:MATH:SMOoth:POINts <integer>
:TRACe:MATH:SMOoth:POINts?
Example: TRAC:MATH:SMO:POIN 501
Notes: Only odd values allowed; if <integer> even, add 1 unless <integer> =
number of sweep points, in which case subtract 1
Used with the TRACe:MATH:SMOoth command.
Preset: 11
Min: 3
Max: Number of sweep points
Initial S/W Revision: Prior to A.02.00