ACP Measurement
Meas Setup
Rel Lim (Car)
Enters a relative limit value for the carrier level. This sets the amplitude levels to test against for the
specified offsets.
The amplitude level is relative to the carrier amplitude. If multiple offsets are available, the list contains
six (6) entries. The offset closest to the carrier channel is the first one in the list.
[:SENSe]:ACP:OFFSet:LIST:TEST selects the type of testing to be done at each offset.
[:SENSe]:ACP:OFFSet[n]:LIST[n]:TEST selects the type of testing to be done at each offset.
You can turn off (not use) specific offsets with the [:SENSe]:ACP:OFFSet[n]:LIST[n]:STATe command.
The query returns the six (6) sets of real numbers that are the current amplitude test limits, relative to the
carrier, for each offset.
Offset[n] n = 1 is base station and n = 2 is mobiles. The default is base station (1).
Key Path Meas Setup, Offset/Limits, Limits,
Mode WCDMA, C2K, 1xEVDO, LTE
Remote Command [:SENSe]:ACPower:OFFSet[1]|2:LIST:RCARrier <real>,
<real>, <real>, <real>, <real>, <real>
[:SENSe]:ACPower:OFFSet[1]|2:LIST:RCARrier?
Example ACP:OFFS2:LIST:RCAR 0,0,0,0,0,0
ACP:OFFS2:LIST:RCAR?
Notes Offset sub op code. 1 for BTS, 2 for MS. Default is BTS.
You must be in the mode that includes ACP measurements to use this
command. Use :INSTrument:SELect to set the mode.
Couplings None
Preset WCDMA: –44.2, –49.2, –49.2, –49.2, –49.2, –49.2|–32.2, –42.2, –42.2, –42.2,
–42.2, –42.2
C2K: 0, 0, 0, 0, 0, 0|0, 0, 0, 0, 0, 0
1xEVDO: –45, –55, –55, –55, –55, –55|–45, –55, –55, –55, –55, –55
LTE: –44.2,–44.2,–44.2,–44.2,–44.2,–44.2|–29.2, –29.2, –29.2, –29.2, –29.2,
–29.2
State Saved Saved in instrument state.
Min –150
Max 50.0
Backwards Compatibility SCPI [:SENSe]:MCPower:OFFSet[1]|2:LIST:RCARrier (PSA WCDMA)
Initial S/W Revision Prior to A.02.00
Modified at S/W Revision A.02.00, A.03.00, A.04.00