Agilent Technologies U9072A cdma-2000 Blood Glucose Meter User Manual


  Open as PDF
of 1403
 
419
ACP Measurement
Meas Setup
Remote Command [:SENSe]:ACPower:OFFSet[1]|2:LIST:BANDwidth[:INTegratio
n] <bandwidth>, <bandwidth>, <bandwidth>, <bandwidth>,
<bandwidth>, <bandwidth>
[:SENSe]:ACPower:OFFSet[1]|2:LIST:BANDwidth[:INTegratio
n]?
Example ACP:OFFS2:LIST:BAND 2 MHz, 2 MHz, 2 MHz, 2 MHz, 2 MHz, 2 MHz
ACP:OFFS2:LIST:BAND?
Notes When setting these values remotely, the position in the list sent corresponds to
the offset. Missing values are not permitted i.e. if you want to change values 2
you must send all values up to 2. Subsequent values will remain unchanged,
unless the number of values sent is greater than the number of carriers, then
subsequent values will be ignored.
Offset sub op code. 1 for BTS, 2 for MS. Default is BTS.
You must be in the mode that includes ACP measurements to use this
command. Use :INSTrument:SELect to set the mode.
Couplings Changing Integ BW might affect to the Span. See Span section for details.
Preset WCDMA: 3.84 MHz, 3.84 MHz, 3.84 MHz, 3.84 MHz, 3.84 MHz, 3.84
MHz|3.84 MHz, 3.84 MHz, 3.84 MHz, 3.84 MHz, 3.84 MHz, 3.84 MHz
WIMAX OFDMA: 10 MHz, 10 MHz, 10 MHz, 10 MHz, 10 MHz, 10 MHz|
10 MHz, 10 MHz, 10 MHz, 10 MHz, 10 MHz, 10 MHz
C2K: 30 kHz, 30 kHz, 30 kHz, 30 kHz, 30 kHz, 30 kHz| 30 kHz, 30 kHz,
30 kHz, 30 kHz, 30 kHz, 30 kHz
TD-SCDMA: 1.28 MHz, 1.28 MHz, 1.28 MHz, 1.28 MHz, 1.28 MHz,
1.28 MHz
1xEVDO: C2K: 30 kHz, 30 kHz, 30 kHz, 30 kHz, 30 kHz, 30 kHz| 30 kHz,
30 kHz, 30 kHz, 30 kHz, 30 kHz, 30 kHz
LTE, LTETDD: 4.515 MHz, 4.515 MHz, 4.515 MHz, 4.515 MHz, 4.515
MHz, 4.515 MHz|4.5 MHz, 4.5 MHz, 4.5 MHz, 4.5 MHz, 4.5 MHz, 4.5 MHz
State Saved Saved in instrument state.
Min 10 Hz
Max 1 GHz
Backwards Compatibility SCPI [:SENSe]:ACPower:OFFSet[1]|2:LIST:BWIDth[:INTegration]
[:SENSe]:ACPR:OFFSet[1]|2:LIST:BANDwidth
[:SENSe]:ACPR:OFFSet[1]|2:LIST:BWIDth
[:SENSe]:MCPower:OFFSet[1]|2:LIST:BANDwidth[:INTegration] (PSA
Power Suite)
[:SENSe]:MCPower:OFFSet[1]|2:LIST:BWIDth[:INTegration] (PSA Power
Suite)
Initial S/W Revision Prior to A.02.00