Spectrum Emission Mask Measurement
Meas Setup
leakage effect to the offset power integration.
Video BW
Changes the analyzer post–detection filter.
Key Path: Meas Setup, Offset/Limits
Mode: WCDMA, C2K, WIMAX OFDMA, TD-SCDMA, 1xEVDO, LTE, LTETDD,
WLAN
Remote Command: [:SENSe]:SEMask:OFFSet[1]|2:LIST:BANDwidth:IMULti
<integer>, …
[:SENSe]:SEMask:OFFSet[1]|2:LIST:BANDwidth:IMULti?
Example: SEM:OFFS2:LIST:BAND:IMUL 1,1,1,1,1,1
SEM:OFFS2:LIST:BAND:IMUL?
Notes: Comma separated list of values.
OFFSet1 is for BTS, 2 for MS. Default is BTS.
You must be in the mode that includes SEM measurement to use this
command. Use :INSTrument:SELect to set the mode.
Couplings: This parameter must adhere to the rule (N x Res BW) <= (Stop freq of the
offset - Start freq of the offset), where N is the multiplier. If the Res Bw is
changed, the multiplier will be changed to ensure this.
Preset: For modes (except WLAN), the preset value is as follows.
WCDMA: 1, 1, 1, 10, 1, 1|1, 1, 1, 1, 1, 1
C2K: 10, 1, 1, 1, 1, 1|1, 1, 1, 1, 1, 1
WIMAX OFDMA, 1xEVDO: 1, 1, 1, 1, 1, 1|1, 1, 1, 1, 1, 1
TD-SCDMA:1, 1, 1, 20, 1, 1|1, 1, 20, 1, 1, 1
LTE, LTETDD: 2, 1, 1, 1, 1, 1|2, 2, 1, 1, 1,1
---------------
WLAN: 1, 1, 1, 1, 1, 1, 1, 1, 1, 1, 1, 1
State Saved: Saved in instrument state.
Min: 1
Max: 1000
Backwards Compatibility SCPI: [:SENSe]:SEMask:OFFSet[1]|2:LIST:BWIDth:IMULti
Initial S/W Revision: Prior to A.02.00
Modified at S/W Revision: A.02.00, A.03.00, A.10.00, A.11.00
Key Path:
Meas Setup, Offset/Limits