Sweep/Control
Gate Delay Compensation
This function allows you to select an RBW-dependent value by which to adjust the gate delay, to
compensate for changes in the delay caused by RBW effects.
Youcan select between uncompensated operation and two types of compensation,
Delay Until RBW
Settled
and Compensate for RBW Group Delay.
See “More Information” on page 1293
Example: SWE:EGAT:HOLD 0.0002
SWE:EGAT:HOLD?
SWE:EGAT:HOLD:AUTO ON
SWE:EGAT:HOLD:AUTO?
Couplings: When
Gate Holdoff is Auto, the Gate Holdoff key shows the value calculated
by the analyzer for the wait time.
Pressing the
Gate Holdoff key while it is in Auto and not selected, causes the
key to become selected and allows the user to adjust the value. If the value is
adjusted, the setting changes to
Man.
Pressing the
Gate Holdoff key, while it is in Auto and selected, does not
change the value of
Gate Holdoff, but causes the setting to change to Man.
Now the user can adjust the value.
Pressing the key while it is in
Man and selected, cause the value to change
back to
Auto.
Pressing the key while it is in
Man and not selected, causes the key to become
selected and allows the user to adjust the value.
When
Method is set to Video or FFT, the Gate Holdoff function has no effect.
Preset: Auto
Auto/On
State Saved: Saved in instrument state
Min: 1 μsec
Max: 1 sec
Initial S/W Revision: Prior to A.02.00
Key Path:
Sweep/Control, Gate
Scope: Meas Global
Remote Command: [:SENSe]:SWEep:EGATe:DELay:COMPensation:TYPE
OFF|SETTled|GDELay
[:SENSe]:SWEep:EGATe:DELay:COMPensation:TYPE?
Example: SWE:EGAT:DEL:COMP:TYPE SETT
SWE:EGAT:DEL:COMP:TYPE?