ACP Measurement
Meas Setup
Positive Offset Limit
Enables you to set the upper limit for the upper segment of the specified offset pair.
Negative Offset Limit
Enables you to set the upper limit for the lower segment of the specified offset pair.
Mode WCDMA, C2K, WIMAX OFDMA, TD-SCDMA, 1xEVDO, LTE, LTETDD
Remote Command :CALCulate:ACPower:OFFSet:LIST:LIMit:POSitive[:UPPer]:D
ATA <real>, <real>, <real>, <real>, <real>, <real>
:CALCulate:ACPower:OFFSet:LIST:LIMit:POSitive[:UPPer]:D
ATA?
Example CALC:ACP:OFFS:LIST:LIM:POS:DATA 0,0,0,0,0,0
CALC:ACP:OFFS:LIST:LIM:POS:DATA?
Notes SCPI only command
Preset WCDMA: –44.2, –49.2, –49.2, –49.2, –49.2, –49.2|–32.2, –42.2, –42.2, –42.2,
–42.2, –42.2
C2K: 0, 0, 0, 0, 0, 0|0, 0, 0, 0, 0, 0
1xEVDO: –45, –55, –55, –55, –55, –55|–45, –55, –55, –55, –55, –55
State Saved Saved in instrument state.
Min –150.0
Max 50.0
Backwards Compatibility SCPI :CALCulate:MCPower:OFFSet:LIST:LIMit:POSitive[:UPPer]:DATA (PSA
Power Suite)
Initial S/W Revision A.04.00
Mode WCDMA, C2K, 1xEVDO, LTE
Remote Command :CALCulate:ACPower:OFFSet:LIST:LIMit:NEGative[:UPPer]:D
ATA <real>, <real>, <real>, <real>, <real>, <real>
:CALCulate:ACPower:OFFSet:LIST:LIMit:NEGative[:UPPer]:D
ATA?
Example CALC:ACP:OFFS:LIST:LIM:NEG:DATA 0,0,0,0,0,0
CALC:ACP:OFFS:LIST:LIM:NEG:DATA?
Notes SCPI only command
Preset WCDMA: –44.2, –49.2, –49.2, –49.2, –49.2, –49.2|–32.2, –42.2, –42.2, –42.2,
–42.2, –42.2
C2K: 0, 0, 0, 0, 0, 0|0, 0, 0, 0, 0, 0
1xEVDO: –45, –55, –55, –55, –55, –55|–45, –55, –55, –55, –55, –55