Trigger
RF path.
Key Path: Trigger, RF Burst
Scope: Meas Global
Remote Command: :TRIGger[:SEQuence]:RFBurst:LEVel:ABSolute <ampl>
:TRIGger[:SEQuence]:RFBurst:LEVel:ABSolute?
Example: TRIG:RFB:LEV:ABS 10 dBm
sets the trigger level of the RF burst envelope signal to the absolute level of 10
dBm
Notes: Sending this command does not switch the setting from relative to absolute; to
switch it you need to send the :TRIGger[:SEQuence]:RFBurst:LEVel:TYPE
command, below.
Amplitude Corrections are not taken into account by the Absolute Trigger
Level. For example, if you have given yourself effective gain with an
amplitude correction factor, the Absolute Trigger will not fire until you have
set the trigger level that far below the displayed signal level, rather than
simply to the displayed signal level. This is only true for Amplitude
Corrections, not External Gain or Ref Level Offset functions.
If mode is Bluetooth, the default value is –50 dBm.
Couplings: This same level is used for the RF Burst trigger source in the Trigger menu,
for the RF Burst selection in the Periodic Timer sync source (in the Trigger
menu and in the Gate Source menu), and also for the RF Burst selection in the
Gate Source menu
Preset: –20 dBm
State Saved: Saved in instrument state
Min: –200 dBm
Max: 100 dBm
Default Unit: depends on the current selected Y-Axis unit
Backwards Compatibility SCPI: :TRIGger[:SEQuence]:FRAMe:RFBurst:LEVel:ABSolute
Initial S/W Revision: Prior to A.02.00
Modified at S/W Revision: A.04.00
Key Path:
Trigger, RF Burst
Remote Command: :TRIGger[:SEQuence]:RFBurst:LEVel:TYPE
ABSolute|RELative
:TRIGger[:SEQuence]:RFBurst:LEVel:TYPE?