Spectrum Emission Mask Measurement
View/Display
View/Display
Accesses a menu of functions that enable you to control the instrument display.
The following keys select how the results are displayed:
Abs Pwr Freq-displays the absolute power levels in dBm and the corresponding frequencies in the
text window.
Rel Pwr Freq-displays the relative power levels in dBc and the corresponding frequencies in the text
window.
Integrated Power-displays the absolute and relative power levels integrated throughout the
bandwidths between the start and stop frequencies in the text window.
“View Selection by Name (Remote Command Only)” on page 566
“Views Selection by Number (Remote Command only)” on page 566
View Selection by Name (Remote Command Only)
Views Selection by Number (Remote Command only)
The following numerical selections determine how the results are displayed:
1. displays the absolute power levels in dBm and the corresponding frequencies in the text window.
2. displays the relative power levels in dBc and the corresponding frequencies in the text window.
Key Path: View/Display
Mode: WCDMA, C2K, WIMAX OFDMA, TD-SCDMA, 1xEVDO, LTE, LTETDD,
WLAN
Remote Command: :DISPlay:SEMask:VIEW[:SELect]
APFReq|RPFReq|IPOWer|CINFormation
:DISPlay:SEMask:VIEW[:SELect]?
Example: DISP:SEM:VIEW IPOW
DISP:SEM:VIEW?
Notes: You must be in the mode that includes SEM measurement to use this
command. Use :INSTrument:SELect to set the mode.
Preset: WCDMA, C2K, TD-SCDMA, 1xEVDO, LTE, LTETDD: APFReq
WIMAX OFDMA, WLAN: RPFReq
State Saved: Saved in instrument state.
Range: Abs Pwr & Freq| Rel Pwr & Freq|Integrated Power|Carrier Info
Initial S/W Revision: Prior to A.02.00
Modified at S/W Revision: A.02.00, A.03.00, A.10.00