1012
Marker Function
Marker Function Off
Turns off band functions for the selected marker.
Band Adjust
Opens a menu that lets you set the width or left or right edges of the band.
It is legal to change the width of the band even if there is no marker function on. Generally this can only
happen by sending the SCPI command since access to the menu is restricted if no marker function is on.
Band/Interval Span
Sets the width of the span for the selected marker.
It is legal to change the width of the band even if there is no marker function on. Generally this can only
happen by sending the SCPI command since access to the menu is restricted if no marker function is on.
In the table below, sweep_width = max(1,sweep_points–1) and sweep_points is the number of sweep
points, set in the
Sweep menu.
Key Path Marker Function
Example :CALC:MARK:FUNC OFF turns off marker functions for marker 1
Notes See the description under the “Marker” on page 989
key, above.
Dependencies Fixed markers: It is not possible to change the Band Function for a Fixed
marker; so all of the Band Function keys are grayed out for a Fixed marker,
including Off
Couplings Turning off the marker function has no effect on the band span nor does it turn
the marker off.
Initial S/W Revision Prior to A.02.00
Key Path
Marker Function
Dependencies If the marker is Fixed, Band Adjust is grayed out.
If the marker function is Off, Band Adjust is grayed out.
Couplings If any of the Band Adjust functions are the active function, the wings and
arms of the selected marker display in green; otherwise they display in white.
Initial S/W Revision Prior to A.02.00
Key Path
Marker Function, Band Adjust
Remote Command :CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:
BAND:SPAN <freq>
:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:FUNCtion:
BAND:SPAN?