Spectrum Emission Mask Measurement
Meas Setup
Abs Stop
Sets the absolute power level limit at the stop frequency for the selected offset. The absolute power level
limit ranges from –200 to +50 dBm. You can also toggle this function between couple and manual. If set
to Couple, the
Abs Stop power level limit is coupled to Abs Start to result in a flat limit line. If set to
Man, Abs Start and Abs Stop take different values to result in a sloped limit line.
The SCPI query returns the five (5) sets of real values currently set to the offset stop absolute power
limits.
Modified at S/W Revision: A.02.00. A.03.00, A.10.00, A.11.00
Key Path:
Meas Setup, Offset/Limits, Limits
Mode: WCDMA, C2K, WIMAX OFDMA, TD-SCDMA, 1xEVDO, LTE, LTETDD,
WLAN
Remote Command: [:SENSe]:SEMask:OFFSet[1]|2:LIST:STOP:ABSolute <real>, …
[:SENSe]:SEMask:OFFSet[1]|2:LIST:STOP:ABSolute?
[:SENSe]:SEMask:OFFSet[1]|2:LIST:STOP:ABSolute:COUPle
ON|OFF|1|0, …
[:SENSe]:SEMask:OFFSet[1]|2:LIST:STOP:ABSolute:COUPle?
Example: SEM:OFFS:LIST:STOP:ABS –12.50 dBm, –24.50 dBm, –24.50 dBm, –11.50
dBm, –11.50 dBm, –11.50 dBm
SEM:OFFS1:LIST:STOP:ABS?
SEM:OFFS:LIST:STOP:ABS:COUP ON, OFF, ON, ON, ON, ON
SEM:OFFS:LIST:STOP:ABS:COUP?
Notes: Comma separated list of values.
OFFSet1 is for BTS, 2 for MS. Default is BTS.
You must be in the mode that includes SEM measurement to use this
command. Use :INSTrument:SELect to set the mode.
Couplings: Coupled to Abs Start if "Auto" is selected, that is, the Stop value is equal to
the Start value.