Input/Output
Capture Buffer
Some WCDMA and demod measurements support this feature. This allows sharing of the raw data
across certain measurements. If you want to make another measurement on the same signal, you would
store that raw data using the "Current Meas -> Capture Buffer" key. Then the data is available for the
next measurement to use. You must have raw data stored in the instrument memory before the Capture
Buffer choice is available for use.
Current Meas -> Capture Buffer
Pressing this key stores the raw data of one measurement in the internal memory of the instrument where
it can then be used by a different measurement by pressing "Stored Data". When raw data is stored, then
the data source selection switch automatically changes to "Stored Data". Stored raw data cannot be
directly accessed by a user. There is no save/recall function to save the raw data in an external media.
However if you want to get the stored raw data, you must first perform a measurement using the stored
raw data. Now you can access the used raw data, which is the same as stored raw data, using the FETch
or READ commands.
Corrections
This key accesses the Amplitude Corrections menu.
Readback: Inputs
Initial S/W Revision: Prior to A.02.00
Key Path:
Input/Output, Data Source
Example: FEED:DATA STOR causes stored measurement data to be used with a
different measurement that supports this.
Notes: Does not auto return. This key is grayed out when you switch to a
measurement that does not support this feature.
Dependencies: If you switch to a measurement that does not support this feature, then the
instrument switches to use "Inputs" and grays out this key. If the grayed out
key is pressed, it generates a message.
Readback: Stored Data
Initial S/W Revision: Prior to A.02.00
Key Path:
Input/Output, Data Source
Remote Command: [:SENSe]:FEED:DATA:STORe
Example: FEED:DATA:STOR stores recorded data
Notes: This is command only, there is no query
Backwards Compatibility SCPI: [:SENSe]:FEED:SOURce:STORe
Initial S/W Revision: Prior to A.02.00