Corrections files are text files in .csv (comma separated values) form, to make them importable into
Excel or other spreadsheet programs. The format for Corrections files is as follows.
Line # Type of field Example Notes
1 File type, must be
“Amplitude Correction”
Amplitude Correction May not be omitted
2 File Description (in quotes) “Correction Factors for
11966E”
60 characters max; may be empty but
may not be omitted. If exceeds 60
characters, error –233 Too much data
reported
3 Comment (in quotes) “Class B Radiated” 60 characters max; may be empty but
may not be omitted. . If exceeds 60
characters, error –233 Too much data
reported
4 Instrument Version, Model
#
A.02.06,N9020A May be empty but may not be omitted
5 Option List, File Format
Versi on
K03 LFE EXM ,01 May be empty but may not be omitted
6 Freq Unit to be used for all
frequency values in the file
Frequency Unit,MHz assumed to be Hz if omitted
7 Antenna Unit Antenna Unit,None If omitted leaves the Antenna unit
unchanged. The amplitude unit in the
Antenna Unit field is a conversion
factor that is used to adjust the Y Axis
Units of the current mode, if the mode
supports Antenna Units. For more
details on antenna correction data, refer
to the Input/Output,Corrections key
description. Allowable values:
dBuv/m, dBuA/m, dBG, dBpT, None
8 Freq Interpolation Frequency
Interpolation,Linear
if omitted leaves the Freq Interpolation
unchanged. Allowable values: Linear,
Logarithmic
9 Bias value in mA Bias,0.00 If omitted leaves the Bias value
unchanged (added as of A.08.50)
10 Bias State Bias State,On If omitted leaves the Bias State
unchanged. Allowable values: On, Off
(added as of A.08.50)
11 Overlap, two values, Freq1
and Freq2, separated by
commas.
Overlap,33500,40000 Uses Freq Unit from line 6. Thus, in
this example Freq1=33.5 GHz, Freq2=
40.0 GHz (see note below). If omitted
leaves the overlap unchanged (added as
of A.08.50)
12 DATA marker DATA Corrections data begins in the next line