System Functions
System
Multiport Adapter Cables Test
This key gives you access to diagnostic capabilities for the RF and trigger cable connections between
EXT and Multiport Adapter with this instrument, which include:
EXT RF Input <-> Multiport Adapter RF OUT
EXT RF Output <-> Multiport Adapter RF IN
EXT RF IO2 <-> Multiport Adapter GPS IN
EXT Trigger 1<-> Multiport Adapter Trigger 1
EXT Trigger 2 <-> Multiport Adapter Trigger 2
Quick Test …
This key gives you access to launch a Windows program for conducting a confidence check of the
Agilent EXT Wireless Communications Test Set and the E6617A Multiport Adapter connected to EXT.
The operator must exit Quick Test to return to the instrument application.
Example: SYST:TEST:WCTS:FEC:RES?
Notes: The output is an IEEE Block format of the Show FEC Self Test Results
contents. Each line is separated with a new-line character.
Initial S/W Revision: A.12.50
Key Path:
System, Diagnostics
Remote Command: :CALibration:MPADapter:CABLes:TEST
Example:
:CAL:MPAD:CABL:TEST
Notes: If the Multiport Adapter cables are not connected correctly. It will report the
proper error, for example:
“–330, Self-test failed, MPA’S RF IN or RF OUT not properly connected”.
Dependencies: This key does not appear unless a multiport adapter is plugged in to the USB.
Grayout error: –221.1400; Multiport Adapter Not Available
Initial S/W Revision: A.10.0
Key Path:
System, Diagnostics
Notes: Operator is responsible for exiting the Quick Test and returning focus to the
Instrument Application.
The softkey in the menu is only displayed when the instrument is an EXT and
the EXTQuickTest.exe file is present in the C:\Program
Files\Agilent\EXTQuickTest folder.
Initial S/W Revision: A.09.49