Input/Output
An Ampcor array can contain 2000 total points, maximum.
Output Config
Accesses keys that configure various output settings, like the frequency reference output, trigger output
and analog output.
Trig Out (1 and 2)
Select the type of output signal that will be output from the rear panel Trig 1 Out or Trig 2 Out
connectors.
Remote Command: [:SENSe]:MPADapter:CORRection:CSET[1]|2|3|4|5|6|7|8|9|1
0|11|12|13|14|15|16:DATA:MERGe <freq>, <ampl>, ...
Example: MPAD:CORR:CSET[1]|2|3|4|5|6|7|8|9|10|11|12|13|14|15|16:DATA:MERGE
15000000,–5.0,25000000,5.0
This adds two correction points at (15 MHz, –5.0 dB) and (25 MHz, 5.0 dB) to
whatever values already exist in correction set 1.
Preset: Empty after Restore Input/Output Defaults. Survives shutdown/restart of
analyzer application (including power cycle)
Min: Freq: 0 Hz
Amptd: –1000 dBm
Max: Freq: 1 THz
Amptd: +1000 dBm
Initial S/W Revision: A.10.00
Key Path:
Input/Output
Backwards Compatibility Notes: 1. In ESA there was not a user interface to enable the Video Output (Analog
Output), Trigger Output, or Gate Output. In the X-Series each of these
physical connectors requires configuration, thus the user interface has
been added for X-Series, along with the potential for an output you think
is always on to be switched off.
Initial S/W Revision: Prior to A.02.00
Key Path:
Input/Output, Output Config
Remote Command: :TRIGger|TRIGger1|TRIGger2[:SEQuence]:OUTPut
HSWP|MEASuring|MAIN|GATE|GTRigger|OEVen|SPOint|SSWeep|S
SETtled|S1Marker|S2Marker|S3Marker|S4Marker|OFF
:TRIGger|TRIGger1|TRIGger2[:SEQuence]:OUTPut?
Example: TRIG:OUTP HSWP
TRIG2:OUTP GATE