271
System Functions
System
The “Align 20 Hz to 3.6 GHz required” Error Condition will appear when this alignment has expired.
User is now responsible to perform the Align Now, 20 Hz to 3.6 GHz in order to keep the receiver in
warranted operation. This alignment can only be performed by user as it is not part of the Auto Align
process.
Key Path: System, Alignments, RF Preselector, Align Now
Mode: All
Remote Command: :CALibration:RFPSelector:FULL
:CALibration:RFPSelector:FULL?
Example: :CAL:RFPS:FULL
Notes: :CALibration:RFPSelector:FULL? Return 0 if successful
:CALibration:RFPSelector:FULL? Return 1 if failed
When Align 20 Hz to 3.6 GHz is performing the alignment, bit 0 in the Status
Operation register is set. Completion, or termination, will clear bit 0 in the
Status Operation register.
This command is sequential; it must complete before further SCPI commands
are processed. Interrupting the alignment from remote is
accomplished by invoking Device Clear followed by the :ABORt
command. Successful completion will clear bit 1, bit 2 in the Status
Questionable Calibration Extended Needed register and bit 0, bit 1 in Status
Questionable Calibration Extended Failure register.
A failure encountered during alignment will set the Error Condition “20 Hz to
3.6 GHz Alignment Failure” and set bit1, bit 2 in the Status Questionable
Calibration Extended Needed register and bit 9 in Status Questionable
Calibration register.
For model N9038A only.
Dependencies: This key does not appear in other than N9038A models, setting or querying
the SCPI will generate an error.
Couplings: Initializes the time for the Last Align Conducted Now, Conducted Time.
Initializes the time for the Last Align Radiated Now, Radiated Time.
Records the temperature for the Last Align Conducted Now, Conducted
Temperature.
Records the temperature for the Last Align Radiated Now, Radiated
Temperature.
Status Bits/OPC Dependencies: Bit 8 or 9 may be set in the Status Questionable Calibration register.
Bit 1 and 2 may be set in the Status Questionable Calibration Extended
Needed register.
Bit 0 and 1 may be set in the Status Questionable Calibration Extended Failure
register.
Initial S/W Revision: A.08.00