Sweep/Control
Sweep Type Rules
Selects which set of rules will be used for automatically choosing the Sweep Type when Sweep Type is
in Auto.
Auto
This selection is automatically chosen when Auto Couple is pressed. When in Auto, the Sweep Type
Rules are set to Best Dynamic Range. It seems like a very simple Auto function but the use of this
construct allows a consistent statement about what the Auto Couple key does.
Dependencies: When a CISPR detector (Quasi Peak, EMI Average, RMS Average) is selected
for any active trace, the FFT key is grayed out.
When the RF Preselector is on, the FFT key is grayed out.
When Source Mode is set to Tracking, Manual FFT is grayed out.
When Signal ID is on, Manual FFT is grayed out.
Grayed out while in Gated LO (meaning Gate is ON and Gate Method is LO).
Grayed out while in Gated Video (meaning Gate is ON and Gate Method is
Video).
State Saved: Saved in instrument state
Readback: FFT
Initial S/W Revision: Prior to A.02.00
Key Path:
Sweep/Control, Sweep Setup
Remote Command: [:SENSe]:SWEep:TYPE:AUTO:RULes SPEed|DRANge
[:SENSe]:SWEep:TYPE:AUTO:RULes?
Dependencies: In Zero Span, this key is irrelevant and cannot be accessed (because the whole
Sweep Setup menu is grayed out in Zero Span), however its settings can be
changed remotely with no error indication.
Preset: DRANge
State Saved: Saved in instrument state
Backwards Compatibility Notes: The legacy parameter DYNamicrange is unsupported
Initial S/W Revision: Prior to A.02.00
Key Path:
Sweep/Control, Sweep Setup, Sweep Type Rules
Remote Command: [:SENSe]:SWEep:TYPE:AUTO:RULes:AUTO[:STATe] OFF|ON|0|1
[:SENSe]:SWEep:TYPE:AUTO:RULes:AUTO[:STATe]?
Example: :SWE:TYPE:AUTO:RUL:AUTO ON
Couplings: Pressing Auto Couple always sets Sweep Type Rules to Auto.