Meas Setup
Delete All Limits
Deletes all limit lines. Pressing Delete All Limits purges the data from all limit line tables.
All limit data are cleared and returned to factory preset settings.
When this key is pressed a prompt is placed on the screen that says “Please press Enter or OK key to
delete all limits. Press ESC or Cancel to close this dialog.” The deletion is only performed if you press
OK or Enter; if so, after the deletion, the informational message “All Limits deleted” appears in the MSG
line.
Limit Line Data (Remote Command Only, Backwards Compatibility)
Defines the limit line values, and destroys all existing data. Up to 200 points may be defined for each
limit using the following parameters.
<x>Frequency or time values as specified by :Calculate:LLINe:CONTrol:DOMain. Units default to Hz
(for frequency) and seconds (for time).
Range: –30 Gs to +30 Gs for time limits, –3 kHz to +350 GHz for frequency limits.
<ampl>Amplitude values units default to dBm. Up to two amplitude values can be provided for each
x-axis value, by repeating <x-axis> in the data list.
Range: –1000 dBm to +1000 dBm
<connect> connect values are either "0" or "1." A "1" means this point is connected to the previously
defined point to define the limit line. A "0" means that it is a point of discontinuity and is not connected
to the preceding point. The connect value is ignored for the first point.
Key Path Meas Setup, Limits
Remote Command :CALCulate:LLINe:ALL:DELete
Example :CALC:LLIN:ALL:DEL deletes all data for all limit lines.
Initial S/W Revision A.02.00
Remote Command: :CALCulate:LLINe[1]|2|3|4|5|6:DATA <x>,<ampl>,<connect>
:CALCulate:LLINe[1]|2|3|4|5|6:DATA?
Example: :CALC:LLIN3:DATA
1E9,–20,0,2E9,–20,1,2E9,–10,1,3E9,–10,1
describes a stair-stepped limit line.
Preset: Limit line data is cleared by Restore Mode Defaults. However, it survives
shutdown/restart of the test set application (including power cycle)
State Saved: Saved in instrument state.
Initial S/W Revision: A.02.00