System Functions
System
Calibrated
Sets the Timebase DAC to the value established during factory or field calibration. The value displayed
on the menu key is the calibrated value.
User
Allows setting the Timebase DAC to a value other than the value established during the factory or field
calibration. The value displayed on the menu key is the calibrated value.
Notes: If the value of the timebase is changed the alignment system automatically
performs an alignment or alerts that an alignment is due.
If the value of the timebase is changed the alignment system automatically
performs an alignment or alerts that an alignment is due.
Preset: This is unaffected by Preset but is set to CALibrated on a “Restore System
Defaults->Align”.
State Saved: No
Initial S/W Revision: Prior to A.02.00
Key Path:
System, Alignments, Timebase DAC
Mode: All
Example: :CAL:FREQ:REF:MODE CAL
Readback Text: [xxx] < where xxx is the calibrated value
Initial S/W Revision: Prior to A.02.00
Key Path:
System, Alignments, Timebase DAC
Mode: All
Example: :CAL:FREQ:REF:MODE USER
Readback Text: xxx < where xxx is the Timebase DAC setting
Initial S/W Revision: Prior to A.02.00
Key Path:
System, Alignments, Timebase DAC
Mode: All
Remote Command: :CALibration:FREQuency:REFerence:FINE <integer>
:CALibration:FREQuency:REFerence:FINE?
Example: :CAL:FREQ:REF:FINE 8191
Notes: If the value of the timebase is changed the alignment system automatically
performs an alignment or alerts that an alignment is due.
Couplings: Setting :CAL:FREQ:REF:FINE sets :CAL:FREQ:REF:MODE USER