Input/Output
Sweeping (HSWP)
Selects the Sweeping Trigger signal to be output to the Trig 1 Out or Trig 2 Out connector when a
measurement is made. This signal has historically been known as "HSWP" (High = Sweeping), and is 5
V TTL level with 50 ohm output impedance."
Measuring
Selects the Measuring trigger signal to be output to the Trig 1 Out or Trig 2 Out connector. This signal is
true while the Measuring status bit is true.
Main Trigger
Selects the current instrument trigger signal to be output to the Trig 1 Out or Trig 2 Out connector.
Gate Trigger
Selects the gate trigger signal to be output to the Trig 1 Out or Trig 2 Out connector. This is the source of
the gate timing, not the actual gate signal.
Gate
Selects the gate signal to be output to the Trig 1 Out or Trig 2 Out connector. The gate signal has been
delayed and its length determined by delay and length settings. When the polarity is positive, a high on
Key Path: Input/Output, Output Config, Trig 1/2 Output
Example: TRIG1:OUTP HSWP
Readback: Sweeping
Initial S/W Revision: Prior to A.02.00
Key Path:
Input/Output, Output Config, Trig 1/2 Output
Example: TRIG1:OUTP MEAS
Readback: Measuring
Initial S/W Revision: Prior to A.02.00
Key Path:
Input/Output, Output Config, Trig 1/2 Output
Example: TRIG1:OUTP MAIN
Readback: Main Trigger
Initial S/W Revision: Prior to A.02.00
Key Path:
Input/Output, Output Config, Trig 1/2 Output
Example: TRIG1:OUTP GTR
Readback: Gate Trigger
Initial S/W Revision: Prior to A.02.00