Channel Power Measurement
Marker
Marker X Axis Value (Remote Command Only)
Sets the marker X Axis value in the current marker X Axis Scale unit. It has no effect if the control mode
is
Off, but is the SCPI equivalent of entering an X value if the control mode is Normal, Delta, or Fixed.
Marker X Axis Position (Remote Command Only)
Sets the marker X Axis Scale position in trace points. This setting has no effect if the control mode is
Off, but is the SCPI equivalent of entering a value if the control mode is Normal or Delta . The entered
value is immediately translated into the current X Axis Scale units for setting the value of the marker.
State Saved: Saved in instrument state.
Range: Normal|Delta|Off
Initial S/W Revision: Prior to A.02.00
Modified at S/W Revision: A.02.00, A.03.00
Mode: WCDMA, C2K, WIMAX OFDMA, 1xEVDO, LTE, LTETDD, WLAN
Remote Command: :CALCulate:CHPower:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:X
<real>
:CALCulate:CHPower:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:X
?
Example: CALC:CHP:MARK3:X 0
CALC:CHP:MARK3:X?
Notes: The query returns the marker’s absolute X Axis value if the control mode is
Normal, or the offset from the marker’s reference marker if the control mode
is
Delta. The query is returned in the fundamental units for the current marker
X Axis scale: Hz for
Frequency.
Preset: After a preset, all markers are turned OFF, so Marker X Axis Value query
returns a not a number (NAN).
State Saved: Saved in instrument state.
Min: –9.9E+37
Max: 9.9E+37
Initial S/W Revision: Prior to A.02.00
Modified at S/W Revision: A.02.00, A.03.00
Mode: WCDMA, C2K, WIMAX OFDMA, 1xEVDO, LTE, LTETDD, WLAN
Remote Command: :CALCulate:CHPower:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:X
:POSition <real>
:CALCulate:CHPower:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:X
:POSition?