Frequency Interpolation
This setting controls how the correction values per-bucket are calculated. We interpolate between frequencies in
either the logarithmic or linear scale.
This setting is handled and stored individually per correction set.
See “Interpolation” on page 963
Interpolation
For each bucket processed by the application, all of the correction factors at the frequency of interest (center
frequency of each bucket) are summed and added to the amplitude. All trace operations and post processing treat
this post-summation value as the true signal to use.
To effect this correction, the goal, for any particular start and stop frequency, is to build a correction trace, whose
number of points matches the current Sweep Points setting of the instrument, which will be used to apply
corrections on a bucket by bucket basis to the data traces.
For amplitudes that lie between two user specified frequency points, we interpolate to determine the amplitude
value. You may select either linear or logarithmic interpolation between the frequencies.
If we interpolate on a log scale, we assume that the line between the two points is a straight line on the log scale.
For example, let’s say the two points are (2,4) and (20,1). A straight line between them on a log scale looks like:
Readback: "None"
Initial S/W Revision: A.10.00
Key Path:
Input/Output, RF Output & Test Set Config, Multiport Adapter,
Corrections, Properties
Remote Command: [:SENSe]:MPADapter:CORRection:CSET[1]|2|3|4|5|6|7|8|9|1
0|11|12|13|14|15|16:X:SPACing LINear|LOGarithmic
[:SENSe]:MPADapter:CORRection:CSET[1]|2|3|4|5|6|7|8|9|1
0|11|12|13|14|15|16:X:SPACing?
Example: :MPAD:CORR:CSET:X:SPAC LIN
Dependencies: This menu selection does not have any effect unless Input/Output, More, RF
Output & Test Set Config, Multiport Adapter, Multiport Adapter is set to the
On state.
Preset: Unaffected by a Preset. Set to Linear by Restore Input/Output Defaults.
State Saved: Saved in instrument state.
Initial S/W Revision: A.10.00