Agilent Technologies U9072A cdma-2000 Blood Glucose Meter User Manual


  Open as PDF
of 1403
 
11 79
Source (Internal)
Query ARB Memory File List (Remote Command Only)
Queries the test set for the list of waveform segments in the ARB memory.
NOTE This command returns a string for waveform segment names in ARB memory. If you want
a string list of waveform segments in the ARB memory, use “Query ARB Memory Full
File List (Remote Command Only)” on page 1179.
Query ARB Memory Full File List (Remote Command Only)
Queries the test set for the string list of waveform segments in the ARB memory. It returns a string list for
waveform segment names in the ARB memory.
Notes: When in Sequence Analyzer mode and Include Source is Yes, an attempt to
delete all files from ARB memory is rejected with an error.
When Include Source is No and you attempt to delete all files from ARB
memory when the ARB is currently playing a file, all files except the one
playing are deleted and an error is generated.
If you attempt to delete all files from ARB memory when there are waveform
files used in “List Sequencer” on page 1219 and “Sequencer” on page 1219
state is ON, all files except the files currently being used in list sequencer are
deleted, and an error is generated.
If the ARB is ON and you load a file to ARB memory or delete a file from
ARB memory, the playing waveform segment may not keep phase continuity
during the ARB memory operation. The waveform will be replayed after the
ARB operation is finished.
Initial S/W Revision: A.05.00
Remote Command: :SOURce:RADio:ARB:CATalog?
Example: SOUR:RAD:ARB:CATalog?
Notes: The return data is in the following format:
<integer> - memory used
<integer> - memory free
<string> … - comma separated list of waveform segments within ARB
memory
Initial S/W Revision: A.05.00
Remote Command: :SOURce:RADio:ARB:FCATalog?
Example: SOUR:RAD:ARB:FCATalog?