System
NOTE This function is NOT available on EXT E6607A and E6607B models.
FEC Self Test
This key invokes the EXT E6607C froned end control self test. When operation is complete, the
generated test summary information is appended to log file E:\Agilent\Instrument\FECTestLog.txt. This
test summary file can be retrieved from the instrument using the MMEM set of SCPI command, once
you have the fully qualified the path and file name.
If the self test fails, the following error message will be generated:
“–330, Self-test failed, see log file E:\Agilent\Instrument\FECTestLog.txt”
If the self test passes, an advisory message “FEC self-test completed successfully” is generated.
NOTE This function is NOT available on EXT E6607A and E6607B models.
Show Result
This key gives you access to show results of the following self tests:
“Source Self Test Results” on page 297
E6607C embeded MPA self-test results - “RFIO Self Test REsults” on page 299
E6607C FEC self-test results - “FEC Self Test REsults” on page 300
Source Self Test Results
Provides a display of last source test results, the display should appear listing model number, serial number and test
time at the top of display, and then list test date/time, test name, measured value, valid range and pass/fail of each
Notes: Access log with command :
MMEM:DATA? "E:\ Agilent\Instrument\RFIOTestLog.txt"
Initial S/W Revision: A.12.50
Key Path:
System, Diagnostics, Self Test
Remote Command: :SYSTem:TEST:WCTS:FEC
Example: SYST:TEST:WCTS:FEC
Notes: Access log with command :
MMEM:DATA? "E:\ Agilent\Instrument\FECTestLog.txt"
Initial S/W Revision: A.12.50
Key Path:
System, Diagnostics, Self Test
Initial S/W Revision: A.12.50