System Functions
System
All Self Test
This key invokes all the self tests defined in the Diagnostics Self Test section.
Source Self Test
This key invokes the internal source self test. When operation is complete, the generated test summary
file is: E:\Agilent\Instrument\CRFSSelfTestLog.txt. This test summary file can be retrieved from the
instrument using the MMEM set of SCPI command, once you have the fully qualified the path and file
name.
If the self test fails, the following error message will be generated:
“–330, Self-test failed, see log file E:\Agilent\Instrument\ CRFSSelfTestLog.txt”
If the self test passes, an advisory message “Source self-test completed successfully” is generated.
RFIO Self Test
This key invokes the RFIO ports self test for embedded MPA, which just are available on E6607C. When
operation is completed, the generated test summary information is appended to log file
E:\Agilent\Instrument\RFIOTestLog.txt. This test summary file can be retrieved from the instrument
using the MMEM set of SCPI command, once you have the fully qualified the path and file name.
If the self test fails, the following error message will be generated:
“–330, Self-test failed, see log file E:\Agilent\Instrument\RFIOTestLog.txt”
If the self test passes, an advisory message “RFIO self-test completed successfully” is generated.
Key Path: System, Diagnostics, Self Test
Remote Command: :SYSTem:TEST:WCTS:[ALL]
Example: SYST:TEST:WCTS:[ALL]
Initial S/W Revision: A.12.50
Key Path:
System, Diagnostics, Self Test
Remote Command: :SYSTem:TEST:WCTS:SOURce
Example: SYST:TEST:WCTS:SOURce
Notes: Access log with command : MMEM:DATA? "E:\
Agilent\Instrument\CRFSSelfTestLog.txt"
Alias of source self test (:SOURce:SELF:TEST[:ALL])
Initial S/W Revision: A.10.00
Key Path:
System, Diagnostics, Self Test
Remote Command: :SYSTem:TEST:WCTS:MPADapter
Example: SYST:TEST:WCTS:MPAD