365
Channel Power Measurement
Trace/Detector
Trace/Detector
Accesses a menu of functions that enable you to control the detectors for the current measurement.
Trace Type
Allows you to select the type of trace you want to use for the current measurement. The first page of this
menu contains a 1–of–N selection of the trace type (
Clear Write, Average, Max Hold, Min Hold) for the
selected trace.
Detector
Accesses a menu of functions that enable you to control the detectors for the current measurement. The
following choices are available:
Key Path: Front-panel key
Initial S/W Revision: Prior to A.02.00
Key Path:
Trace/Detector
Mode: WCDMA, C2K, WIMAX OFDMA, 1xEVDO, LTE, LTETDD, WLAN
Remote Command: :TRACe:CHPower:TYPE WRITe|AVERage|MAXHold|MINHold
:TRACe:CHPower:TYPE?
Example: TRAC:CHP:TYPE WRIT
TRAC:CHP:TYPE?
Notes: WRITe = Clear Write
AVERage = Average
MAXHold = Maximum Hold
MINHold = Minimum Hold
Couplings: When Detector setting is “Auto” ([:SENSe]:CHPower:DETector:AUTO?),
Detector ([:SENSe]:CHPower:DETector[:FUNCtion]?) switches aligning
with the switch of this parameter: “NORMal” with WRITe (Clear Write),
“AVERage” with AVERage, “POSitive (peak)” with MAXHold, and
“NEGative (peak)” with MINHold.
Preset: AVERage
State Saved: Saved in instrument state.
Range: ClearWrite|Average|MaxHold|MinHold
Initial S/W Revision: Prior to A.02.00
Modified at S/W Revision: A.02.00, A.03.00