Multiport Adapter Amplitude Correction
This key selects the Multiport Adapter Amplitude Corrections as the data type to be imported. When
pressed a second time, it brings up the Select Menu, which lets you select the Correction into which the
data will be imported.
Amplitude Corrections are fully discussed in the documentation of the Input/Output key, under the
Corrections softkey.
A set of preloaded Corrections files can be found in the directory
/My Documents/ EMC Limits and Ampcor.
Under this directory, the directory called Ampcor (Legacy Naming) contains a set of legacy corrections
files, generally the same files that were supplied with older Agilent EMI analyzers, that use the legacy
suffixes .ant, .oth, .usr, and .cbl, and the old 8-character file names. In the directory called Ampcor, the
same files can be found, with the same suffixes, but with longer, more descriptive filenames.
When the Amplitude Correction is an Antenna correction and the Antenna Unit in the file is not
None,
the Y Axis Unit setting will change to match the Antenna Unit in the file.
Initial S/W Revision: A.02.00
Key Path:
Recall, Data, Amplitude Correction
Mode: SA|EDGEGSM|PN
Remote Command: :MMEMory:LOAD:MPADapter:CORRection
1|2|3|4|5|6|7|8|9|10|11|12|13|14|15|16, <filename>.
Example: :MMEM:LOAD:MPAD:CORR 2 "myAmpcor.csv" recalls the Multiport
Adapter Amplitude Correction data from the file myAmpcor.csv in the current
directory to the 2nd Multiport Adapter Amplitude Correction table, and turns
on MPA Correction 2.
The default path is My Documents\amplitudeCorrections.