Front-Panel Features (E6607C)
6 MPA RFIO, Ports 0 - 7 RF input and output connections to the DUTs (SMA connectors).
The maximum safe input level at any of these ports is 2 W (+33 dBm), +
15 Vdc.
7 GPS, Ports 0 - 3 RF input and output connections to the DUTs (SMA connectors).
The maximum safe input level at any of these ports is 0.25 W (+24 dBm), 0 Vdc.
Because GPS Ports 0 - 3 are the outputs of a four-way splitter, the maximum output
power levels from these ports are lower than for RFIO Ports 0 - 7; see the E6607C
data sheet for specifics.
8 RF OUT GPS RF output port (communicates only with the source).
The maximum safe input level at any of these ports is 0.25 W (+24 dBm), 0 Vdc.
9 DIST IN GPS RF distribution input port (SMA connector). This port is connected to the
four-way splitter that provides the signal to the 4 GPS output Ports 0 - 3.
The maximum safe input level at any of these ports is 0.25 W (+24 dBm), 0 Vdc.
10 RF OUT Tx/DL RF output port (communicates only with the source).
The maximum safe input level at any of these ports is 0.25 W (+24 dBm), +
15 Vdc.
11 MPA IN Tx/DL MPA RF input port (SMA connector).
The maximum safe input level at any of these ports is 0.25 W (+24 dBm), +
15 Vdc.
12 RF IN Rx/UL RF input port (communicates only with the analyzer).
The maximum safe input level at any of these ports is 0.25 W (+24 dBm), +
15 Vdc.
13 MPA OUT Rx/UL RF MPA output port (SMA connector).
The maximum safe input level at any of these ports is 0.25 W (+24 dBm), +
15 Vdc.