If self test fails, error message “–330, Self-test failed, see log file E:\Agilent\Instrument\
CRFSSelfTestLog.txt” is generated. If self test passes, an advisory message “Source self-test completed
successfully” is generated.
Recall
Most of the functions under this key work the same way in many measurements, so they are documented
in the System Functions section. For details about this key, see “Recall” on page 192
Save
Most of the functions under this key work the same way in many measurements, so they are documented
in the Utility Functions section. For details about this key, see “Save” on page 204
Signal Studio Commands
Overview
EXT supports connectivity with Signal Studio. To achieve this, the SCPI commands described in this
chapter enable you to connect Signal Studio to the EXT test set and to download waveform into the hard
disk. They are SCPI only, no menu
Query License List (Remote Command Only)
This query provides a listing of the current licenses for external software installed for the test set internal
source.
Query License List Detail (Remote Command Only)
This is an obsolete command for Signal Studio. However, it has not been removed from Signal Studio to
prevent a Signal Studio connectivity time out. Using this command returns a null string and does not
Key Path: System, Diagnostics, Internal Source, Self Test
Remote Command: :SOURce:SELF:TEST[:ALL]
Example: SOUR:SELF:TEST
Notes: MMEM:DATA? "E:\ Agilent\Instrument\CRFSSelfTestLog.txt"
Initial S/W Revision: A.09.20
Initial S/W Revision: A.05.00
Remote Command: :SYSTem:LICense:EXTernal:LIST?
Example: SYST:LIC:EXT:LIST?
Notes: Query only.
Initial S/W Revision: A.05.00