507
Spectrum Emission Mask Measurement
Meas Setup
Sweep Time
Sets the sweep time used to calculate the power in the reference channel. Sweep Time can be set
manually or put in auto mode.
Min: 1 kHz
Max: 645 MHz
Initial S/W Revision: Prior to A.02.00
Modified at S/W Revision: A.02.00, A.03.00
Key Path:
Meas Setup, Ref Channel
Mode: WCDMA, C2K, WIMAX OFDMA, TD-SCDMA, 1xEVDO, LTE, LTETDD,
WLAN
Remote Command: [:SENSe]:SEMask:SWEep[1]|2:TIME <time>
[:SENSe]:SEMask:SWEep[1]|2:TIME?
[:SENSe]:SEMask:SWEep[1]|2:TIME:AUTO OFF|0|ON|1
[:SENSe]:SEMask:SWEep[1]|2:TIME:AUTO?
Example: SEM:SWE:TIME 9ms
SEM:SWE:TIME?
SEM:SWE:TIME:AUTO OFF
SEM:SWE:TIME:AUTO?
Notes: Sweep Time sub op code, 1 is for BTS, 2 for MS. Default is BTS.
You must be in the mode that includes SEM measurement to use this
command. Use :INSTrument:SELect to set the mode.
Couplings: When the Sweep Time is set manually, Auto is set to OFF.
Value is coupled with Channel Detector selection, Channel Resolution BW,
Channel Video BW if the state is Auto.
When set to Auto, the Sweep Time is automatically calculated
Preset: Automatically calculated
ON
State Saved: Saved in instrument state.
Min: 1 ms
Max: 4000 s
Initial S/W Revision: Prior to A.02.00
Modified at S/W Revision: A.02.00, A.03.00