Trigger
SECAM-L
Sets the TV standard to SECAM-L.
Auto/Holdoff
Opens up a menu that lets you adjust Auto Trigger and Trigger Holdoff parameters
Auto Trig
Sets the time that the analyzer will wait for the trigger conditions to be met. If they are not met after that
much time, then the analyzer is triggered anyway.
Readback: PAL-N
Initial S/W Revision: Prior to A.02.00
Key Path:
Trigger, TV, Standard
Example: TRIG:TV:STAN LSEC
Readback: SECAM-L
Initial S/W Revision: Prior to A.02.00
Key Path:
Trigger
Readback line: Displays a summary of the Auto Trig and Holdoff settings, in square brackets
First line: Auto Off or Auto On
Second Line: “Hldf” followed by:
• If Holdoff is Off, readback Off
• If Holdoff On and Type = Normal, readback value
• If Holdoff On and Type = Above, readback value followed by AL
• If Holdoff On and Type = Below, readback value followed by BL
• If Holdoff Type selection is not supported by the current measurement,
Holdoff Type is always Normal
Initial S/W Revision: A.02.00
Key Path:
Trigger, Auto/Holdoff
Remote Command: :TRIGger[:SEQuence]:ATRigger <time>
:TRIGger[:SEQuence]:ATRigger?
:TRIGger[:SEQuence]:ATRigger:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:ATRigger:STATe?
Example: TRIG:ATR:STAT ON
TRIG:ATR 100 ms