Reset + Run
Sets Reset and Run as the trigger response for the continuous trigger type. Reset and Run sets the waveform
generator to play a waveform sequence or segment continuously when the first trigger is received. Subsequent
triggers reset the waveform sequence or segment to the start, and then play it continuously.
Single
Sets the active trigger type to Single. If Single is already selected as the active trigger type, pressing this key allows
access to the single trigger type setup menu. In Single trigger mode, the waveform plays once.
No Retrigger
Selects No Retrigger as the trigger response for single trigger type. No Retrigger sets the waveform generator to
play a waveform sequence or segment once when a trigger is received. Any triggers then received during playback
are ignored.
Example: SOUR:RAD:ARB:TRIG:TYPE:CONT TRIG
Initial S/W Revision: A.05.00
Key Path:
Source, Modulation Setup, ARB, Trigger Type, Continuous
Example: SOUR:RAD:ARB:TRIG:TYPE:CONT RES
Initial S/W Revision: A.05.00
Key Path:
Source, Modulation Setup, ARB, Trigger Type
Remote Command: :SOURce:RADio:ARB:RETRigger ON|OFF|IMMediate
:SOURce:RADio:ARB:RETRigger?
Example: SOUR:RAD:ARB:RETR OFF
Notes: ON: Buffered Trigger
OFF: No Retrigger
Immediate: Restart on Trigger
This is defined as an enumerated SCPI command, with ON|OFF being
considered as enumerated types rather than Boolean. This means the query
will return OFF instead of 0, and ON instead of 1.
Preset: ON
Range: No Retrigger | Buffered Trigger | Restart on Trigger
Initial S/W Revision: A.05.00
Key Path:
Source, Modulation Setup, ARB, Trigger Type, Single
Example: SOUR:RAD:ARB:RETR OFF
Initial S/W Revision: A.05.00