Mode
Application Catalog Revision
Returns the Revision of the provided Model Number.
Application Catalog Options
Returns a list of Options for the provided Model Number
Notes: Returned value is a quoted string of a comma separated list of Model
Numbers. Example, if SAMS and Phase Noise are installed and licensed:
"N9060A,N9068A"
String length is COUNt * 7 – 1. (7 = Model Number length + 1 for comma. –1
= no comma for the 1st entry.)
Preset: Not affected by a Preset
State Saved: Not saved in instrument state.
Initial S/W Revision: Prior to A.02.00
Remote Command: :SYSTem:APPLication:CATalog:REVision? <model>
Example: :SYST:APPL:CAT:REV? ‘N9060A’
Notes: Returned value is a quoted string of revision for the provided Model Number.
The revision will be a null-string ("") if the provided Model Number is not
installed and licensed. Example, if SAMS is installed and licensed:
"1.0.0.0"
Preset: Not affected by a Preset.
State Saved: Not saved in instrument state.
Initial S/W Revision: Prior to A.02.00
Remote Command: :SYSTem:APPLication:CATalog:OPTion? <model>
Example: :SYST:APPL:CAT:OPT? ‘N9060A’
Notes: Returned value is a quoted string of a comma separated list of Options, in the
same format as *OPT? or :SYSTem:OPTion?. If the provided Model Number
is not installed and licensed a null-string ("") will be returned. Example, if
SAMS is installed and licensed:
"2FP"
String length is a maximum of 255 characters.
Preset: Not affected by a Preset
State Saved: Not saved in instrument state.
Initial S/W Revision: Prior to A.02.00