11 35
Single (Single Measurement/Sweep)
Single (Single Measurement/Sweep)
Sets the test set for Single measurement operation. The single/continuous state is Meas Global, so the
setting will affect all the measurements. If you are Paused, pressing
Single does a Resume.
Key Path: Front-panel key
Example: :INIT:CONT OFF
Notes: See
Cont key description.
Backwards Compatibility Notes: For Spectrum Analysis mode in ESA and PSA, the
Single hardkey and the
INITiate:IMM switched from continuous measurement to single measurement
and restarted sweeps and averages (displayed average count reset to 1), but
did not restart
Max Hold and Min Hold. In the X-Series, the Single hardkey
and the INITiate:IMM command initiate a sweep/ measurement/ average
sequence/hold sequence including
Max Hold and Min Hold.
For Spectrum Analysis mode in ESA and PSA, the
Single hardkey restarted
the sweep regardless of whether or not you were in an active sweep or sweep
sequence. In the X-Series,
Restart does this but Single only restarts the
sweep or sweep sequence if you are in the idle state.
Initial S/W Revision: Prior to A.02.00