Agilent Technologies U9072A cdma-2000 Blood Glucose Meter User Manual


  Open as PDF
of 1403
 
11 42
Source (Internal)
ALC, or when the modulation consists of slow amplitude variations that the automatic leveling would
remove.
Modulation
Allows you to toggle the state of the modulation.
Frequency
Allows you to access the Frequency sub-menu.
Key Path: Source, Amplitude
Remote Command: :SOURce:POWer:ALC[:STATe] ON|OFF|1|0
:SOURce:POWer:ALC[:STATe]?
Example: SOUR:POW:ALC OFF
SOUR:POW:ALC?
Preset: On
Range: On | Off
Initial S/W Revision: A.05.00
Key Path:
Source
Remote Command: :OUTPut:MODulation[:STATe] ON|OFF|1|0
:OUTPut:MODulation[:STATe]?
Example: :OUTP:MOD OFF
Notes: This setting is for independent mode and has no effect on “List Sequencer” on
page 1219. If the “Sequencer” on page 1219 is set to ON, the list sequencer
controls the source output and this key will be grayed-out. And this setting
will be none-forceful grey out on front panel to indicate out-of-scope.
Non-forceful means user still can change this setting by SCPI but cannot
change manually on front panel. When setto Off will make source leave list
sequencer and this setting will be black out and take effect immediately When
the Modulation is ON, the “MOD” annunciator is displayed in the system
settings panel. When the Modulation is turned Off, the “MOD” annunciator is
cleared. If the “Sequencer” on page 1219 is set to ON, the “MOD”
annunciator will be replaced by “SEQ” in the system settings panel indicating
that the output is controlled by list sequencer.
Preset: Off
Range: On | Off
Initial S/W Revision: A.05.00
Key Path:
Source