Meas Setup
trace does not turn red if it crosses a secondary limit line.
Type
Selects whether the limit you are editing is an upper or lower limit. An upper limit fails if the trace
exceeds the limit. A lower limit fails if the trace falls below the limit.
Key Path Meas Setup, Limits, Properties
Remote Command :CALCulate:LLINe[1]|2|3|4|5|6:TRACe 1|2|3|4|5|6
:CALCulate:LLINe[1]|2|3|4|5|6:TRACe?
Example :CALC:LLIN3:TRAC 2 applies limit 3 to trace 2.
Notes When the trace display is off, the trace is not tested. The trace is tested only
when the trace display is on and Test Limits (see “Test Limits” on page 1064)
is on.
Couplings This matters when testing a trace or limit line for failure, via
:CALC:LLIN3:FAIL? or :CALC:TRAC2:FAIL?
Preset Limits 1 and 2 preset to 1, Limits 3 and 4 preset to 2, Limits 5 and 6 preset to 3
Not affected by Mode Preset, preset by Restore Mode Defaults.
State Saved Saved in State.
Min 1
Max 6
Readback Trace 1|2|3|4|5|6
Initial S/W Revision A.02.00
Key Path
Meas Setup, Limits, Properties
Remote Command :CALCulate:LLINe[1]|2|3|4|5|6:TYPE UPPer|LOWer
:CALCulate:LLINe[1]|2|3|4|5|6:TYPE?
Example :CALC:LLIN2:TYPE LOW sets limit line 2 to act as a lower limit.
Couplings If a margin has already been set for this limit line, and this key is used to
change the limit type, then the margin value will reverse sign.
Preset Upper for Line 1, 3, and 5; Lower for Line 2, 4, 6.
Not affected by Mode Preset, preset by Restore Mode Defaults.
State Saved Saved in State.
Initial S/W Revision A.02.00