Meas Setup
Limit Clear (Remote Command Only, SCPI standard Compatibility)
Clears a limit line, and all associated data. This command is identical to “:CALC:LLIN:DEL”
Trace Fail? (Remote Command Only)
Tests a trace against all associated limit lines. Returns a 0 if the trace is within all limits and margins, a 1
if the trace exceed either the limit or the margin. If no limits apply to the selected trace, this will
automatically return a 0.
Only applies to limits that are turned on, if a Limit is off it will not be tested. If a Trace is not displaying
it will still be tested, and if
Test Limits (All Limits) is off the Trace will still be tested.
This command ignores limit lines that are assigned to other traces.
Fixed / Relative Limit (Remote Command Only, Backwards Compatibility)
This command sets both Relative to CF and Relative to RL simultaneously for all limits. If queried, it
returns whether Limit Line 1 is set Relative to CF, and ignores all other fixed/relative data.
Remote Command: :CALCulate:LIMit[1]|2|3|4|5|6:CLEar
Example: :CALC:LIM2:CLE deletes all data for limit line 2.
Couplings: This command is identical to :CALC:LLIN:DEL
Initial S/W Revision: A.02.00
Remote Command: :CALCulate:TRACe[1]|2|3|4|5|6:FAIL?
Example: :CALC:TRAC3:FAIL? returns a zero if there is no failure, 1 if the trace
exceeds either the limit or the margin.
Initial S/W Revision: A.02.00
Remote Command: :CALCulate:LLINe:CMODe FIXed|RELative
:CALCulate:LLINe:CMODe?
Example: :CALC:LLIN:CMOD REL makes all limit lines relative to the center
frequency and reference level.