Input/Output
corrections. Press ESC or Cancel to close this dialog.” The deletion is only performed if you press OK or Enter.
Set (Replace) Data (Remote Command Only)
The command takes an ASCII series of alternating frequency and amplitude points, each value separated by
commas.
The values sent in the command will totally replace all existing correction points in the specified set.
An Ampcor array can contain 2000 points maximum.
Merge Correction Data (Remote Command Only)
The command takes an ASCII series of alternating frequency and amplitude points, each value separated by
commas. The difference between this command and Set Data is that this merges new correction points into an
existing set.
Any new point with the same frequency as an existing correction point will replace the existing point’s amplitude
with that of the new point.
Key Path:
Input/Output, RF Output & Test Set Config, Multiport Adapter,
Corrections
Remote Command: [:SENSe]:MPADapter:CORRection:CSET:ALL:DELete
Example: MPAD:CORR:CSET:ALL:DEL
Initial S/W Revision: A.10.00
Remote Command: [:SENSe]:MPADapter:CORRection:CSET[1]|2|3|4|5|6|7|8|9|1
0|11|12|13|14|15|16:DATA <freq>, <ampl>, . . .
[:SENSe]:MPADapter:CORRection:CSET[1]|2|3|4|5|6|7|8|9|1
0|11|12|13|14|15|16:DATA?
Example: MPAD:CORR:CSET[1]|2|3|4|5|6|7|8|9|10|11|12|13|14|15|16:DATA
10000000,–1.0,20000000,1.0
This defines two correction points at (10 MHz, –1.0 dB) and (20 MHz, 1.0
dB) for correction set 1.
Preset: Empty after Restore Input/Output Defaults. Survives a shutdown or restart of
analyzer application (including a power cycle).
State Saved: Saved in instrument state.
Min: Freq: 0 Hz
Amptd: –1000 dBm
Max: Freq: 1 THz
Amptd: +1000 dBm
Initial S/W Revision: A.10.00