System Functions
System
The example of FEC self test result display is as follows:
Show FEC Self Test Results contents (Remote Command Only)
A remote command is available to obtain the contents of the Show FEC Self Test Results screen (the entire
contents, not just the currently displayed page).
FEC Self Test Results
Produce Number: E6607C
Serial Number: MY51380437
Instrument S/W: 11/16/2012 2:51:19 PM
FpgaVersionTest
Date Time(GMT) Name MeasValue ValidRange Resul
t
11/23/2012 16:13:56 Analog_FPGA 16.000 >= 16.000 Pass
11/23/2012 16:13:56 Digital_FPGA 50.000 >= 46.000 Pass
11/23/2012 16:13:56 CRFS_FPGA 38.000 >= 38.000 Pass
PowerSupplyTest
Date Time(GMT) Name MeasValue ValidRange Resul
t
11/23/2012 16:13:56 ABUS_+32CHK 31.904 30.900 - 32.900 Pass
11/23/2012 16:13:56 ABUS_+12CHK 12.296 10.800 - 13.200 Pass
11/23/2012 16:13:56 +10VA 9.935 9.600 - 10.200 Pass
11/23/2012 16:13:56 +5VA 4.995 4.900 - 5.100 Pass
11/23/2012 16:13:56 +3.3VA 3.299 3.200 - 3.400 Pass
11/23/2012 16:13:56 -3.3VA -3.311 -3.400 - -3.200 Pass
11/23/2012 16:13:56 ACOM 0.00 -0.200 – 0.200 Pass
11/23/2012 16:13:56 -5VA -5.036 -5.100 - -4.900 Pass
11/23/2012 16:13:56 -6.1VA -5.880 -6.200 - -5.700 Pass
11/23/2012 16:13:56 -10VA -10.116 -10.200 - -9.800 Pass
11/23/2012 16:13:56 ABUS_-2.5V_REF -2.508 -2.520 - -2.470 Pass
11/23/2012 16:13:56 ABUS_+2.5V_REF 2.508 2.480 - 2.520 Pass
11/23/2012 16:13:56 ABUS_-10VPALC -10.047 -10.200 - -9.800 Pass
11/23/2012 16:13:57 ABUS_DET_MOD_FLT 18.000 7.800 - 100.000 Pass
Remote Command: :SYSTem:TEST:WCTS:FEC:RESult?