Modulation Accuracy (Composite Rho)
AMPTD Y Scale
AMPTD Y Scale
Accesses the AMPTD Y Scale menu that allows you to set the desired vertical scale and associated
settings. This menu is available when in “I/Q Error” view. It is not available when in the following
views: “I/Q Measured Polar Graph” or “Power, Timing and Phase”.
Ref Value
Sets the absolute power reference value ranging from –500.00 to 500.00 dBm with 0.01 dB resolution.
However, since Auto Scaling defaults to On, this value is automatically determined by the measurement
result. When you set a value manually, Auto Scaling automatically changes to Off.
I/Q Error View - EVM Window
Sets the absolute power reference value in the EVM window.
Key Path Front-panel key
Key Path AMPTD Y Scale
Mode cdma2000
Remote Command :DISPlay:RHO:VIEW2:WINDow[1]:TRACe:Y[:SCALe]:RLEVe
l <real>
:DISPlay:RHO:VIEW2:WINDow[1]:TRACe:Y[:SCALe]:RLEVe
l?
Example DISP:RHO:VIEW2:WIND:TRAC:Y:RLEV 0.0
DISP:RHO:VIEW2:WIND:TRAC:Y:RLEV?
Restriction and Notes If Auto Scaling is set to On, this value is automatically determined by
the measurement result. When you set this value manually, Auto
Scaling is automatically set to Off.
Dependencies/Couplings See Restriction and Notes
Remote Command Notes WINDow[1]: Evm window on I/Q Error view
Preset 0.00
State Saved Saved in instrument state.
Min –500.0
Max 500.0
Key Path
AMPTD Y Scale