ACP Measurement
Meas Setup
Avg Mode
Enables you to set the averaging mode. This determines the averaging action after the specified number
of data acquisitions (average count) is reached.
When set to Exponential (Exp) the measurement averaging continues using the specified number of
averages to compute each averaged value. The average will be displayed at the end of each sweep.
When set to Repeat, the measurement resets the average counter each time the specified number of
averages is reached.
Carrier Setup
Accesses a menu that contains Carriers, Ref Carrier, Ref Car Freq, Ref Car Pwr and Configure Carriers.
Carriers
Specifies the number of carriers to be measured.
Key Path Meas Setup
Mode WCDMA, C2K, WIMAX OFDMA, TD-SCDMA, 1xEVDO, LTE, LTETDD
Remote Command [:SENSe]:ACPower:AVERage:TCONtrol EXPonential|REPeat
[:SENSe]:ACPower:AVERage:TCONtrol?
Example ACP:AVER:TCON EXP
ACP:AVER:TCON?
Notes You must be in the mode that includes ACP measurements to use this
command. Use :INSTrument:SELect to set the mode.
Preset EXPonential
State Saved Saved in instrument state.
Range Exp|Repeat
Backwards Compatibility SCPI [:SENSe]:ACPR:AVERage:TCONtrol
Initial S/W Revision Prior to A.02.00
Modified at S/W Revision A.02.00, A.03.00
Key Path
Meas Setup
Initial S/W Revision Prior to A.02.00
Key Path
Meas Setup, Carrier Setup, Configure Carriers
Mode WCDMA, C2K, WIMAX OFDMA, 1xEVDO, TD-SCDMA, LTE, LTETDD
Remote Command [:SENSe]:ACPower:CARRier[1]|2:COUNt <integer>
[:SENSe]:ACPower:CARRier[1]|2:COUNt?