399
ACP Measurement
Marker
Couple Markers
When this function is On, moving any marker causes an equal X axis movement of every other marker
which is not
Off. By “equal X axis movement” we mean that we preserve the difference between each
marker’s X axis value (in the fundamental x-axis units of the trace that marker is on) and the X axis value
of the marker being moved (in the same fundamental x-axis units).
Marker All Off
Turns all active markers off.
Couplings This is not affected by Auto Coupling.
Sending the remote command causes the addressed marker to become
selected.
Preset All Markers Off
State Saved Saved in instrument state.
Range 1|2|3
Initial S/W Revision Prior to A.02.00
Modified at S/W Revision A.02.00, A.03.00
Key Path
Marker
Mode WCDMA, C2K, WIMAX OFDMA, TD-SCDMA, 1xEVDO, LTE, LTETDD
Remote Command :CALCulate:ACPower:MARKer:COUPle[:STATe] ON|OFF|1|0
:CALCulate:ACPower:MARKer:COUPle[:STATe]?
Example CALC:ACP:MARK:COUP ON
Dependencies Unavailable when “Meas Method” on page 433 is set to RBW.
Preset OFF
State Saved Saved in instrument state.
Range On|Off
Initial S/W Revision Prior to A.02.00
Modified at S/W Revision A.02.00, A.03.00
Key Path
Marker
Mode WCDMA, C2K, WIMAX OFDMA, TD-SCDMA, 1xEVDO, LTE, LTETDD
Remote Command :CALCulate:ACPower:MARKer:AOFF
Example CALC:ACP:MARK:AOFF
Dependencies Unavailable when “Meas Method” on page 433 is set to RBW.