Agilent Technologies U9072A cdma-2000 Blood Glucose Meter User Manual


  Open as PDF
of 1403
 
679
Code Domain
Marker
reference value of the selected marker appears on the Active Function area.
Active Function Display:
Marker symbol value at I/Q Symbol Polar Vector graph
Marker X-axis value at other graphs
Default Active Function: the active function for the selected marker’s current control mode. If the
current control mode is Off, there is no active function and the active function is turned off.
The marker X axis value entered in the active function area displays the marker value to its full entered
precision.
NOTE When Marker Trace is ‘POLar’ (I/Q Polar), ‘Delta’ is not supported.
Remote Command :CALCulate:CDPower:MARKer[1]|2|3|4|5|6|7|8|9|10|11|1
2:MODE POSition|DELTa|OFF
:CALCulate:CDPower:MARKer[1]|2|3|4|5|6|7|8|9|10|11|1
2:MODE?
Example CALC:CDP:MARK:MODE POS
CALC:CDP:MARK:MODE?
Restriction and Notes If the selected marker is Off, pressing Marker sets it to Normal and
places it at the center of the screen on the trace determined by the
Marker
Trace
rules. At the same time, Marker X Axis Value appears on the
Active Function area.
Default Active Function: the active function for the selected marker’s
current control mode. If the current control mode is Off, there is no active
function and the active function is turned off.
Active Function Display:
Marker symbol value at I/Q Symbol Polar Vector graph
Marker X-axis value at other graphs
the marker X axis value entered in the active function area displays the
marker value to its full entered precision.
Remote Command Notes NORMal is changed to POSition in the new SA.
Preset OFF
State Saved Saved in instrument state.
Range Normal|Delta|Off
Key Path
Marker