Agilent Technologies U9072A cdma-2000 Blood Glucose Meter User Manual


  Open as PDF
of 1403
 
766
Modulation Accuracy (Composite Rho)
SPAN X Scale
I/Q Error View – Mag Error Window
When Auto Scaling is On and the Restart front-panel key is pressed, will result in automatically
displaying the scale per division and reference value results in the magnitude error window.
I/Q Error View – Phase Error Window
When Auto Scaling is On and the Restart front-panel key is pressed, results in automatically displaying
Restriction and Notes When this parameter is set to On, pressing the front-panel Restart key,
or the Restart key in the Meas Control menu, activates the scale
coupling function, that automatically determines scale per division and
reference values based on the measurement results. When you set a
value to either Scale/Div or Ref Value manually, Auto Scaling is
automatically set to Off.
Dependencies/Couplings See Restriction and Notes
Remote Command Notes WINDow[1]: Evm Error window on I/Q Error view
Preset ON
State Saved Saved in instrument state.
Range On|Off
Key Path
Span X Scale
Mode cdma2000
Remote Command :DISPlay:RHO:VIEW2:WINDow2:TRACe:X[:SCALe]:COUPle
0|1|OFF|ON
:DISPlay:RHO:VIEW2:WINDow2:TRACe:X[:SCALe]:COUPle?
Example DISP:RHO:VIEW2:WIND2:TRAC:X:COUP ON
DISP:RHO:VIEW2:WIND2:TRAC:X:COUP?
Restriction and Notes When this parameter is set to On, pressing the front-panel Restart key,
or the Restart key in the Meas Control menu, activates the scale
coupling function, that automatically determines scale per division and
reference values based on the measurement results. When you set a
value to either Scale/Div or Ref Value manually, Auto Scaling is
automatically set to Off.
Dependencies/Couplings See Restriction and Notes
Remote Command Notes WINDow2: Mag Error window on I/Q Error view
Preset ON
State Saved Saved in instrument state.
Range On|Off
Key Path
Span X Scale