Hitachi S-3400N Microscope & Magnifier User Manual


 
6.4.4
6 - 79
(g) Threshold Level combo box
Specifies a threshold value to be taken in the threshold level method. A numeric
value may be input with respect to a difference of "100" between minimum and
maximum in signal variation on the edge part. If "50" is specified, for instance, the
intermediate threshold level is used. This parameter is selectable in a range of 5 to
95.
(10) Apply Button
Applies the form of cursor under display according to the set value. Also saves the
registration No. of the measurement conditions.
(11) Save Button
Saves the present settings under a different registration No. A window for specifying the
registration No. of the save destination appears.
(12) Del Button
Deletes the registration.
(13) Close Button
Closes the Measurement Mode window. When measurement parameters have been