3.5.7
3 - 91
(2) Observable range
The following page shows the X and Y observable ranges by specimen size, and mutual
restrictions on the Z and Tilt axes at the Standard height, and the movable range for the Tilt
axis as a function of specimen height by specimen size. With regard to Z and Tilt
restrictions, for an actual movable range, if either the Z or T scrollbar is set to the
coordinates to which the axis is moved by using the Z/TILT box of the Stage tab on the
Operation Panel, the movable range for the other axis is automatically calculated and
displayed on the scrollbar. Therefore, SOMETHING (subject missing) should be set within
that range.
See 3.5.7.6 Z and Tilt Axis Operation.
The Z axis coodinate value indicated in the Z scroll bar and the Z coodinate box on the
Z/TILT Window of the Stage tab changes according to setting of the specimen height.
Refer to < 3.2.4.2 Measuring and Setting the Specimen Height >.
Fig. 3.5-64 Z/TILT Window
Z scrollbar
Z coordinate box