3.9.8
3 - 133
(5) Recording entry data with images
According to entry data, recording refer to < 3.6.7 Recording Data Display with Images>.
NOTICE: The Auto Data Display mode saves the data displayed on the screen together with
the image itself if the [Embed into Image] checkbox is turned ON in the Record tab of
the Setup window. If the checkbox is turned OFF, the displayed data, even when
appearing on the screen, will not be saved as embedded data.
If the Input Data and Auto Data options are on at the time an image is captured, only
the image can be saved by unchecking the Embed into Image checkbox on the
Captured Image window.
If the data being displayed is to be saved together with the image, turn ON the
checkbox.
NOTICE: If an image is displayed with dimensions, and if the magnification factor is changed
after the image is input, the dimensions are recalculated and re-displayed at the
following timing:
• When the image is frozen
• When Direct Photo operation is performed
• When the image being scanned is fixed, such as executing the Capture function
NOTICE: Dimension marks with measurement calculate dimension using present magnification
value. The dimension value will be re-calculated when image magnification is
changed.
In Dual Screen mode, it is possible to display two different magnification images.
Note that in such a case, dimension value is correct for only one of two images.
An abnormal text character display where text is shown doubly may occur. Although
this will be very rare, carry out the following operation in such case. This will recover
the normal display.