Hitachi S-3400N Microscope & Magnifier User Manual


 
3.8.3
3 - 113
3.8.3 Withdrawing the Specimen
The following procedures can be employed to withdraw a specimen from the specimen chamber:
(1) On the EVAC panel located on the front side of the main unit, press the AIR button.
Alternatively, click the AIR button located in the upper right section of the Control Panel.
This introduces air into the specimen chamber. Wait approximately 100 seconds until the
specimen chamber is completely at the atmospheric pressure.
EVAC panel Control Panel
Fig. 3.8-2 AIR Button (letting air into the specimen chamber)
(2) Grasp the handle on the front side of the specimen stage. Pull the specimen stage toward
you, and take out the specimen. When removing the specimen, be sure to put on clean
gloves; do not use a bare hand.
Fig. 3.8-3 Pulling Out the Stage Fig. 3.8-4 Removing the Specimen
IMPORTANT: Do not touch the operation parts such as gears and driving screws and note that
clothes are never placed when you bring the hand close to the movement
mechanism in the specimen stage by the specimen exchange etc.