Hitachi S-3400N Microscope & Magnifier User Manual


 
3.5.6
3 - 55
Fig. 3.5-25 Allowable Angle of Tilt (Type I) with respect to the Z Axis
(negative angle of tilt)
IMPORTANT: Set Z and T within the allowable range; otherwise, the specimen may strike the
objective lens or BSE detector and cause damage to both specimen and them.
3.5.6.5 Operating the Z-axis (varying the Working Distance [WD])
Turning the Z-axis knob, counterclockwise moves the specimen toward the objective lens, and
reduces the WD. Conversely, turning the Z-axis knob clockwise moves the specimen away
from the objective lens, and increases the WD. The value of the WD can be read off from the
scale provided on the Z-axis knob.
Readings from the scale are valid when the combined height of the specimen base on which the
specimen is loaded and the specimen base stand is set at the Standard position on the height
gauge.
-25
-20
-15
-10
-5
0
5
0 5 10 15 20 25 30 35 40
Z [mm]
T [deg]
15mm
26mm
32mm
51mm
77mm, 85mm
102mm
127mm
153mm
200mm