For Safe Handling
SAFETY - 20
Precautions on Proper Operation (cont’d)
Transportation and Relocation of Instrument
(a) The transport of the system, which carries considerable risk,
should never be attempted by a customer. Any system transport
should be performed with the assistance and supervision of
Hitachi-approved engineers.
(b) Do not lift the instrument by holding the table. The strength of
table fitting is not sufficient for bearing the weight of display unit,
approximately 120 kg. Should the table be lifted, the display unit
might slip off and crash. Hence, it is recommended to remove
the table and transport the display unit independently when
moving the instrument.
Ambient Temperature and Humidity
Even when the system is not running, the room in which the system
is installed should be maintained under the following environmental
conditions:
Room temperature: 15℃ to 30℃
Humidity: 70% or lower (no condensation)
If the system is not used for a long time, the specimen chamber
should be evacuated instead of being exposed to the atmosphere.